AP7000 Multiparameter Probe

 

AP-7000 Sleeve Off 440x264

 

Features

  • DO-EC-pH-ORP-Temp-Depth standard with every probe
  •  6 additional sensor ports that can accept any of our ISE or Optical sensors
  • Built in self cleaning, cleans ALL sensors using novel method
  • Tough metal AquaConn connectors are fitted as standard to enable testing at depths of up to 100m 
  • Fully compatible with Neon telemetry data loggers
  • SDI-12 compatible
  • Modbus (RS-485) output

Overview


The AP-7000 is the ground breaking instrument designed for long-term deployment during unmanned water testing.
Readings can logged with standard data loggers or transmitted in real-time using Neon web-based, telemetry data loggers. Readings can also be displayed on the GPS Aquameter.

There are an impressive 12 different optional electrodes available for the AP-7000. Some are ion selective electrodes and some optical, including; Ammonium & Ammonia, Chloride, Fluoride, Nitrate, Calcium, Turbidity, Chlorophyll, Blue Green Algae, Rhodamine, Fluorescein and Refined Oil.

As an option, the AP-7000 can be supplied with additional SDI-12 and Modbus (RS-485) outputs to allow direct connection to a wide range of data logging and telemetry equipment. This option facilitates direct replacement of existing probes made by other manufacturers.

Technical


AP-7000 Mechanical Specification
Protection ClassIP68 (permanent immersion)
Immersion DepthMin 150mm Max 100m
Operating Temperature-5° C – +50°C
Dimensions (L x Dia)450mm x 77mm
Weight (excluding cable)1350g

AP-7000 Standard Electrode Specification
Standard Parameters
Dissolved OxygenRange0 – 500.0% / 0 – 50.00 mg/L
Resolution0.1% / 0.01mg/L
Accuracy0 – 200%: ± 1% of reading. 200% – 500%: ± 10%
Conductivity (EC)Range0 – 200 mS/cm (0 – 200,000 µS/cm)
Resolution3 Auto-range scales: 0 – 9999 µS/cm, 10.00 – 99.99 mS/cm, 100.0 – 200.0mS/cm
Accuracy± 1% of reading or ± 1µS/cm if greater
TDS*Range0 – 100,000 mg/L (ppm)
Resolution2 Auto-range scales: 0 – 9999mg/L, 10.00 – 100.00g/L
Accuracy± 1% of reading or ± 1mg/L if greater
Resistivity*Range5 Ωcm – 1 MΩcm
Resolution2 Auto-range scales: 5 – 9999 Ωcm, 10.0 – 1000.0 KΩcm
Accuracy± 1% of reading or ± 1 Ωcm if greater
Salinity*Range0 – 70 PSU / 0 – 70.00 ppt (g/Kg)
Resolution0.01 PSU / 0.01 ppt
Accuracy± 1% of reading or ± 0.1 unit if greater
Seawater Specific Gravity*Range0 – 50 σt
Resolution0.1 σt
Accuracy± 1.0 σt
pHRange0 – 14 pH / ± 625mV
Resolution0.01 pH / ± 0.1mV
Accuracy± 0.01 pH / ± 5mV
ORPRange± 2000mV
Resolution0.1mV
Accuracy± 5mV
TemperatureRange-5°C – +50°C (23°F – 122°F)
Resolution0.1° C/F
Accuracy± 0.5° C
Optional Parameters
Depth **Range0 – 100m (0 – 330F)
Resolution0.01m/F
Accuracy± 1%
Ion Specific Electrodes***
AmmoniumRange0 – 1,000mg/L (ppm)
Resolution2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 999.9 mg/L
Accuracy± 10% of reading
Ammonia†Range0 – 1,000mg/L (ppm)
Resolution2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 999.9 mg/L
Accuracy± 10% of reading
ChlorideRange0 – 1,000mg/L (ppm)
Resolution2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 999.9 mg/L
Accuracy± 10% of reading
FlourideRange0 – 1,000mg/L (ppm)
Resolution2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 999.9 mg/L
Accuracy± 10% of reading
NitrateRange0 – 1,000mg/L (ppm)
Resolution2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 999.9 mg/L
Accuracy± 10% of reading
CalciumRange0 – 1,000mg/L (ppm)
Resolution2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 999.9 mg/L
Accuracy± 10% of reading
Optical Electrodes***
TurbidityRange0 – 3000 NTU
Resolution2 Auto-range scales: 0.0 99.9 NTU, 100 – 3000 NTU
Repeatability± 2% of reading
ChlorophyllRange0 – 500 µg/L (ppb)
Resolution0.01 µg/L
Repeatability± 2% of reading
Phycocyanin (Freshwater Blue-Green Algae)Range0 – 300,000 cells/mL
Resolution1 cell/mL
Repeatability± 2% of reading
Phycoerythrin (Marine Blue-Green Algae)Range0 – 200,000 cells/mL
Resolution1 cell/mL
Repeatability± 2% of reading
Rhodamine WT DyeRange0 – 500 µg/L (ppb)
Resolution0.01 µg/L
Repeatability± 5% of reading
Fluorescein DyeRange0 – 500 µg/L (ppb)
Resolution0.01 µg/L
Repeatability± 5% of reading
Refined OilRange0 – 10,000 µg/L (ppb)
Resolution0.1 µg/L
Repeatability± 2% of reading
CDOM/FDOMRange0 – 2,000mg/L (ppm)
Resolution2 Auto-range scales:
0.00 – 99.99 mg/L, 100.0 – 2,000mg/L
Repeatability± 10% of reading or 2ppm (whichever is greater
* Readings calculated from EC and temperature electrode values
**Option on the AP-2000. Comes as standard on the AP-7000
† Ammonium electrode required. Readings calculated from Ammonium, pH and temperature electrode values.

*** In addition to the standard parameters supplied with the base Aquaprobe, the AP-2000 has 1 ISE port and 1 optical port, the AP-5000 has 4 ports that can take any combination of ISE or optical sensors and the AP-7000 has 6 ports that can also take any combination of ISE or optical sensors.

The accuracy figures quoted above represent the equipment’s capability at the calibration points at 25°C. These figures do not take into account errors introduced by variations in the accuracy of calibration solutions and errors beyond the control of the manufacturer that may be introduced by environmental conditions in the field.

Documents